Resolution (mass spectrometry)

In mass spectrometry, resolution is a measure of the ability to distinguish between two peaks of slightly different mass-to-charge ratios (m/z) in a mass spectrum. It is a fundamental performance characteristic of mass spectrometers, determining how well adjacent signals can be separated.

Definition

According to the International Union of Pure and Applied Chemistry (IUPAC), resolution in mass spectrometry is defined as the observed m/z value divided by the smallest difference Δ(m/z) for two ions that can be separated:

$$ R = \frac{m/z}{\Delta(m/z)} $$

A larger value of R indicates better separation of peaks. The m/z value at which the measurement was made, as well as the definition and method of measurement of Δ(m/z), should always be reported.

Measurement Methods

Two principal methods are used to determine Δ(m/z):

Peak Width Definition

Δ(m/z) is taken as the width of a single peak measured at a specified fraction of the maximum peak height. Commonly used fractions are 50% (full width at half maximum, FWHM), 5%, or 0.5%. The FWHM definition is the most widely used standard.

Valley Definition

Δ(m/z) is defined as the separation between two adjacent peaks of equal height such that the valley between them is a specified fraction of the peak height. The most common standard is the 10% valley definition, where the valley between two equal-height peaks is no more than 10% of the peak height. A 5% peak width is technically equivalent to a 10% valley definition for isolated symmetrical peaks.

Resolution vs. Resolving Power

There has been historical controversy regarding the terminology. Three distinct conventions exist in the literature:

  1. Resolution and resolving power are equivalent, both expressed as m/Δm (Meyerson 1975; IUPAC 2013).
  2. Resolution is m/Δm and resolving power is Δm (Price 1991; Todd 1991).
  3. Resolving power is m/Δm and resolution is Δm (Beynon 1978).

The 2013 IUPAC recommendations (Murray et al.) define resolution as (m/z)/Δ(m/z) and resolving power as the "measure of the ability of a mass spectrometer to provide a specified value of mass resolution." The majority of the mass spectrometry community uses resolution as defined by IUPAC.

Typical Values by Instrument Type

Instrument Type Resolving Power (FWHM)
Fourier Transform Ion Cyclotron Resonance (FT-ICR-MS) Up to 1,000,000
Orbitrap Up to 100,000–500,000
High-Resolution Time-of-Flight (HR-TOF) 40,000–60,000
Time-of-Flight (TOF) 10,000
Quadrupole / Ion Trap (ultrazoom mode) ~10,000
Quadrupole / Ion Trap (standard) ~1,000

Importance

High resolution is essential for distinguishing ions with small mass differences, such as isobaric interferences in complex mixtures (e.g., biological, environmental, or food samples). It enables accurate mass determination, unambiguous elemental composition assignment, and reliable quantification. Higher resolution generally comes at the cost of reduced sensitivity and slower acquisition rates, creating a trade-off that must be managed based on the analytical application.

References

  • IUPAC. Compendium of Chemical Terminology, 5th ed. (the "Gold Book"). Online version: "resolution in mass spectroscopy". doi:10.1351/goldbook.R05318
  • Murray, K.K., Boyd, R.K., Eberlin, M.N., Langley, G.J., Li, L., Naito, Y. (2013). "Definitions of terms relating to mass spectrometry". Pure Appl. Chem. 85, 1515–1609. doi:10.1351/PAC-REC-06-04-06
  • Murray, K.K. (2022). "Resolution and Resolving Power in Mass Spectrometry". J. Am. Soc. Mass Spectrom. 33(12), 2342–2347. doi:10.1021/jasms.2c00216