ISO 14155

ISO 14155 is an international standard published by the International Organization for Standardization (ISO) that specifies good clinical practice (GCP) for the design, conduct, recording, and reporting of clinical investigations of medical devices carried out in human subjects. Its primary purpose is to protect the rights, safety, and well-being of human subjects while ensuring the scientific validity and credibility of clinical investigation results.

Scope

ISO 14155 applies to any organization conducting clinical investigations of medical devices in human subjects, including medical device manufacturers sponsoring pre-market clinical investigations for regulatory approval, contract research organizations (CROs), clinical investigation sites and principal investigators, ethics committees and institutional review boards (IRBs), and regulatory authorities assessing clinical evidence in device submissions. It does not apply to in vitro diagnostic medical devices, which are governed separately by ISO 20916.

Editions

The standard has been published in multiple editions:

Edition Publication Year Notes
1st Edition 1996 Initial publication
2nd Edition 2003 —
3rd Edition 2011 Revised and expanded
Amendment 2020 ISO 14155:2020
4th Edition 2026 ISO 14155:2026 (published March 23, 2026)

The 4th edition (ISO 14155:2026) introduced significant changes including mandatory risk management integration aligned with ISO 14971, formal requirements for Clinical Events Committees (CECs) and Data Monitoring Committees (DMCs), an estimand framework aligned with ICH E9(R1), stricter eligibility enforcement, enhanced adverse event reporting with dual categorization, and expanded subject protection requirements including implant card templates.

Key Requirements

Clinical Investigation Plan (CIP)

The CIP is the master document governing the clinical investigation. It must include device identification, justification for the investigation, study objectives and endpoints, study design, risk management plan, eligibility criteria, informed consent procedures, adverse event reporting procedures, data management plans, monitoring plans, and governance structures for independent oversight committees.

Sponsor Responsibilities

The sponsor (typically the manufacturer) retains overall responsibility for the clinical investigation and is responsible for planning and initiation, quality assurance, monitoring, safety reporting, data management and analysis, study termination, and financial disclosure.

Principal Investigator Responsibilities

The principal investigator manages day-to-day conduct at each investigation site, including subject management, protocol compliance, data collection and recording, adverse event documentation, medical care, record retention, and site management.

Adverse Event Reporting

ISO 14155:2026 uses a structured adverse event taxonomy defined in its Annex F, including Adverse Events (AEs), Serious Adverse Events (SAEs), Adverse Device Effects (ADEs), Serious Adverse Device Effects (SADEs), and Device Deficiencies. The 2026 edition introduced a mandatory dual categorization requirement for adverse events linked to device deficiencies, assessing each event on both clinical severity and device causality.

Regulatory Alignment

ISO 14155 is recognized by regulatory authorities worldwide:

  • European Union: EN ISO 14155 is harmonized under the EU Medical Device Regulation (MDR) 2017/745 (Articles 62–82 and Annex XV). Compliance provides a presumption of conformity with relevant MDR clinical investigation requirements.
  • United States: The FDA recognizes ISO 14155 and accepts clinical data collected per the standard for regulatory submissions, alongside 21 CFR Part 812 (Investigational Device Exemptions).
  • Other regions: The standard is referenced or accepted in clinical investigation guidance in Japan, China, Australia, Brazil, and Canada.

Relationship to Other Standards

While ISO 14155 governs medical device clinical investigations, ICH E6 GCP governs pharmaceutical clinical trials. The 2026 edition brought these frameworks closer together through the inclusion of the estimand framework. Key differences include risk management (ISO 14971 for devices vs. ICH Q9 for drugs), endpoint types (device performance vs. drug efficacy), and adverse event categorization methodologies.