High Speed LAN Instrument Protocol (HiSLIP) is a standard network protocol used for controlling test and measurement instruments over Local Area Networks (LAN). It is defined by the IVI Foundation under specification IVI-6.1.
Purpose and Design
HiSLIP was developed to replace the legacy VXI-11 protocol (which relies on ONC RPC/Open Network Computing Remote Procedure Call) with a modern, firewall-friendly, and higher-performance alternative. It is designed to provide a reliable, low-latency communication channel for instrument control commands (typically SCPI - Standard Commands for Programmable Instruments) and high-speed data transfer.
Key Technical Characteristics
- Transport Layer: Uses TCP/IP (Transmission Control Protocol/Internet Protocol) as the underlying transport mechanism.
- Port Assignment: The Internet Assigned Numbers Authority (IANA) has assigned TCP port 4880 for HiSLIP.
- Channel Architecture: Supports multiple simultaneous communication channels over a single TCP connection:
- Control Channel: Used for device management, locking, and sending/receiving SCPI commands.
- Data Channels (Optional): Used for high-throughput bulk data transfer (e.g., waveform data, digitizer records) without blocking the control channel.
- Synchronization: Implements a message-based synchronization mechanism (using
SyncandAsyncmessages) to ensure command/response ordering and to handle Device Clear and Trigger operations deterministically. - Locking Mechanism: Provives mandatory support for shared and exclusive locking (via
Lock/Unlockmessages) to manage multi-client access to a single instrument. - Compatibility: Designed to be semantically compatible with VXI-11 operations (Device Write, Device Read, Device Clear, Trigger, Lock) to simplify migration for driver developers (e.g., VISA driver implementations).
Standardization and Adoption
- Specification: IVI-6.1: High Speed LAN Instrument Protocol (HiSLIP).
- Maintenance: Maintained by the IVI Foundation, a consortium of test and measurement vendors (including Keysight Technologies, National Instruments (now NI, part of Emerson), Rohde & Schwarz, Tektronix, and others).
- LXI Compliance: HiSLIP is the mandatory transport protocol for LXI (LAN eXtensions for Instrumentation) Version 1.4 and later devices (specifically LXI Core 2011 and newer), replacing VXI-11 as the required standard for instrument communication in the LXI framework.
Advantages over VXI-11
- Firewall/NAT Traversal: Uses a single, well-known TCP port (4880), simplifying firewall configuration compared to VXI-11's dynamic RPC port mapping (Portmapper/rpcbind on port 111).
- Performance: Lower protocol overhead and support for pipelined commands and asynchronous data channels result in higher data throughput and lower latency.
- IPv6 Support: Native support for IPv6 addressing.
- Robustness: Improved error recovery and connection state management.